data sheet fxe fcx series field indicator date EDS8-49E jan. 18, 2003 this is a field installation type moving-coil indicator which receives and indicates signal of 4 to 20 ma dc. it is employed when an indicator is to be installed separately from the transmitter. functional specifications measured input: 4 to 20ma dc (2-wire type) internal impedance: 3.5 ? or less scale: 0 to 100% linear scale or 0 to 100% square scale, actual scale note: double scale is unavailable. scale length: 50mm environmental protection: iec ip67 and nema4x explosionproof structure: intrinsic safety baseefa eexiaiict4, t5(pending) flameproofing jis d2g4 case material: aluminum alloy finish color: epoxy urethane double coating silver (case cover is blue-silver) ambient temperature: 40 to +85 c for explosionproof units, ambient temper- ature must be within the limits specified by each standard. ambient humidity: 100% rh or less storage temperature: 40 to +90 c mass: approx. 3.5kg conduit connection: g1/2, 1/2 14npt, pg13.5 or m20 x 1.5 conduit, as specified external terminal: input signal terminal x 2, ground terminal x 1, m4mm screw mounting method: mounting on 50a (2b) pipe or on wall outer dimensions (h x w x d): 92 x 222 x 179mm [performance specifications] accuracy: 1.5 class withstand voltage: 500v ac, 50/60 hz, for 1 minute, to ground insulation resistance: 100m ? or more at 500v dc, to ground specifications scope of delivery code symbols 1 e0 0 0 x f- s t w 10 2 1 345678 1112 9 description b c d a b c k a b c y b note: *10th code b, c is only applicable for 4th codes. scale 0 to 100% linear scale 0 to 100% square scale actual scale explosionproof structure general type (non-explosionproof) jis flameproof (conduit seal type)(*) jis flameproof (cable seal type)(*) baseefa intrinsic safety pipe mounting bracket not provided provided (steel) provided (stainless steel) stainless steel tag plate none yes conduit connection g1/2 1/2-14npt pg13.5 m20 x 1.5 v indicator and pipe mounting bracket (as specified)
fxe information in this catalog is subject to change without notice. printed in japan head office 6-17, sanbancho, chiyoda-ku, tokyo 102-0075, japan http://www.fesys.co.jp/eng sales div. international sales dept. no.1, fuji-machi, hino-city, tokyo, 191-8502 japan phone: 81-42-585-6201, 6202 fax: 81-42-585-6187 http: //www.fic-net.jp/eng caution on safety *before using this product, be sure to read its instruction manual in advance. external connection diagram outline diagram (unit:mm) 184 1mounting pipe jis 50a(2b) u-bolt m8 up down field indicator *"b"conduit connection *cable gland (for flame proof, cable seal type only) clamp(only with flame proof type) "a" conduit connection 38 15 179 60 66 58 4- 9 holes 70 88 55 17 92 164 70 120 92 note: asterisked(*) items are provided only with flameproof(cable seal type) type. 11mm cable should be used. 13 g e g1/2 e 4th code s t v w g1/2 e 1/2-14npt pg13.5 m20 1.5 17 f 16 8 16 8 g 5 4.5 5 details of "a" details of "b" fcx transmitter power supply case 1 field indicator fxe + - ck + ck- + - ck + short bar note: maximum three(3) indicators in series can be connected to the transmitter terminals ck+,ck-. ck- field indicator fxe power supply case 2 fcx transmitter + - ck + ck- + - note: maximum three(3) indicators in series can be connected between the power supply and the transmitter.
|